44 EasyData Guidebook
When EasyData detects a sensor, it automatically loads a calibration for that
sensor. If you wish to edit a sensor’s calibration, you can do so by manually
entering new calibration values or by using either a single-point or two-point
calibration procedure.
Manually Inputting a New Calibration
1. From the EasyData main screen, select , and
then select CH1: (or the appropriate channel).
2. Select
.
3. Select
. A menu is displayed showing the different
calibration methods.
4. Select Manual Entry.
5. The slope for the current calibration is displayed. Press
to erase the displayed value. Type in your new
slope value and select
.